MPEE
Volumetric Length measuring error MPEE
A set of 5 gages has to be measured 3 times with one probing at each end, while the block is being aligned in 7 different directions in space. All measuring results must fulfil the Specification "E".
MPEP
Volumetric probing Error MPEP
A Precision sphere has to be measured with 25 equally distributed probings. P is the range of all radii (sphere form). (Rmax - Rmin = Sphere form).
MPETHP
Volumetric Scanning probing Error MPETHP
A precision Sphere is to be scanned with 4 defined lines. THP is the range of all radii (sphere form-scanning) (Rmax - Rmin = Sphere form)
MF, MS, ML
Fixed multiple-stylus probing system error MF, MS, ML